ISO 17025 calibration covers size setting error, counting efficiency, sampling flow rate error, size resolution, and false count in the actual equipment.
Size Setting Error
The size setting error tells you the deviation from the particle counter’s programmed sizes. Per ISO 21501-4:2018 certification guidelines, the particle size must be ±10% to pass the calibration. The Uncertainty of each particle size will be stated in microns.
Counting Efficiency
This metric will be determined by comparing a particle to a National Institute of Standards and Technology (NIST) traceable particle with higher sensitivity. The ISO 21501-4:2018 standard allows for a tolerance of 50% (± 20%) of the minimum detectable size, and 100% (± 10%) of a size 1.5 – 2 times larger than the minimum detectable channel size. The Uncertainty is stated in percent counting efficiency.
Sampling Flow Rate Error
The sampling flow rate tolerance is ±5% of the stated rate for the unit and refers to the amount of air that flows through the laser block of the particle counter. The flow value uncertainty is stated in liters per minute (LPM).
Size Resolution
The size resolution will be stated with the particle size used in microns and the criteria of less than 15% as stated per ISO 21501-4:2018. The uncertainty will be stated in percent size resolution.
False Count
This is also known as the zero-count test. It is the number of particles detected on the minimum detectable channel in a certain volume of air with a 95% confidence level. We also state the false count rate in accordance with JIS B 9921, which gives a limit of one count per five minutes. Both values are stated as average observed counts in five minutes.