Technical Details

  1100 LD 3100 5100 3350 Boulder Counter

Specifications

Size Range 0.1 - 1.0 μm
Optional ranges up to 5.0 µm
0.3 - 25.0μm 0.5 - 25.0μm 0.3 - 25.0 µm 5.0 - 100.0 um
Channel Sizes Standard 8 channels: 0.10, 0.15, 0.2, 0.25, 0.3, 0.5, 0.7, 1.0μm
Optional 8 channels: 0.10, 0.15, 0.2, 0.25, 0.3, 0.5, 1.0, 5.0 µm
Standard: 0.3, 0.5, 1.0, 3.0, 5.0, 10.0 µm
Standard: 0.3, 0.5, 1.0, 5.0, 10.0, 25.0 µm
8-Channel: 0.3, 0.5, 0.7, 1.0, 3.0, 5.0,7.0,10.0 µm
Standard: 0.5, 0.7, 1.0, 3.0, 5.0, 10.0 µm
Standard: 0.5, 1.0, 3.0, 5.0, 10.0, 25.0 µm
8-Channel: 0.5, 0.7, 1.0, 2.0, 3.0, 5.0, 7.0,10.0 µm
Standard: 0.3, 0.5, 1.0, 3.0, 5.0, 10.0 µm
Optional: 0.3, 0.5, 1.0, 5.0, 10.0, 25.0 µm
Standard: 5.0, 10.0, 25.0, 40.0, 50.0,100.0µm 
Flow Rate 1.0 CFM (28.3 LPM) 100 LPM 28.3 LPM
Counting Efficiency 50% @ 0.1 µm; 100% for particles > 0.15 µm per ISO 21501-4 50% @ 0.3 µm; 100% @ sizes
> 0.45 µm
50% @ 0.5 um; 100% @ sizes
> 0.75 µm
50% @ 0.3 µm; 100% @ sizes > 0.45 µm 50% @ 5.0 um
Concentration Limits 500,000 Particles / ft3 (17,150,000 / m3) @ 5% Coincidence Loss 1,000,000 particles/ft3
(35,300,000/m3) @ 10% coincidence loss
Suitable for ISO classes 1-8 cleanrooms
400,000 particles/ft3 (14,120,000/m3)
@ 10% coincidence loss
Suitable for ISO classes 1-8 cleanrooms
500,000 particles/ft3 @ 5% coincidence loss 
Light Source Long Life Laser Diode Extreme Life Laser Diode Technology with >20-year Mean Time To Failure (MTTF)
Dimensions 14.65"Lx8.0"Wx10.4"H (37x20x26cm) 9.5"Lx8.0"Wx10.2"H (24.13x20.32x25.9cm) 14.65"Lx8.0"Wx10.4"H (37x20x26cm) 9.5"Lx8.0"Wx10.2"H (24.13x20.32x25.9cm)
Weight With battery: 23.55 lbs (10.68 kg)
Without battery: 20.45 lbs (9.28 kg)
With battery: 13.40 lbs (6.08 kg)
Without battery: 12.00 lbs (5.44 kg)
With battery: 26.15 lbs (11.86 kg)
Without battery: 23.05 lbs (10.46 kg)
With battery: 13.40 lbs (6.08 kg)
Without battery: 12.00 lbs (5.44 kg)
Zero Count Level < 1 count / 5 minutes (per JIS B 9921)
Calibration Meets ISO 21501-4 Calibration using NIST Traceable PSL Spheres, DMA and Condensation Particle Counter
Count Modes Automatic, Manual, Beep, Concentration, Cumulative/Differential, Real Time (with LMSNet or LMS Express RT software)
Data Storage 3,000 Sample Records, Rotating Buffer (particle and environmental data with location and time)
Communication Modes USB, Ethernet, RS-485 and Secure Data Transfer (LSDT) to and Storage on USB Flash Drive
Supporting Software LMS Express and Express RT; LMS Pharma
Environmental Sensors Optional (up to 4) temperature/relative humidity, air velocity, differential pressure or others
Touch Screen Display 5.7-inch (14.47 cm) Color TFT Touch Screen Display
Reports Pass/Fail Report Wizard for ISO-14644-1, EU GMP Annex 1 and FS-209 Report Formats
Alarms Integrated alarm buzzer and external alarm connector. Alarms on counts, low battery, sensor failure
Sample Output Internally Filtered to 100% @ 0.1μm, Optional HEPA Filtration for Cooling System Exhaust Internally filtered to HEPA standards (>99.97% @ 0.3μm)
Vacuum Source Internal closed-loop controlled pump. Allows the use of 3/8" ID sample tubing of up to 12 meters (40 feet) @ 50 & 100LPM
Enclosure Stainless Steel
Power 100-240 VAC, 50-60 Hz, External AC/DC Adapter
Battery Optional: Li-Ion, Removable & Rechargeable (typical recharge time 7.5 hours)
Languages English (U.S.), French, German, Italian, Spanish, Chinese, Russian, Japanese, Korean

Environmental Conditions

Operating 50°F to 104°F (10°C to 40°C) / 20% to 95% non-condensing
Storage 14°F to 122°F (-10°C to 50°C) / Up to 98% non-condensing

Accessories

Included Operating Manual on CD; Isokinetic Sample Probe with Tripod and Tubing; Purge Filter; Power Cable and Spare Fuse; LMS XChange and Express Software; Printer; Printer Paper
Optional Printed Operating Manual; Network Adapter; Wireless Network Adapter; Internal Removeable, Rechargeable Battery; Carrying Case with Wheels; Cleanroom Cart; Temperature/Relative Humidity Probe; Air Velocity Probe; Differential Pressure Probe; Filter Scanning Probe; High Pressure Difuser; 32-Port Universal Manifold; LMS Express RT Software; LMSNet Software; Validation Documentation Wireless Network Adapter; Removable Rechargeable Li-Ion Battery; Carrying Case with Wheels; Temperature/Relative Humidity Probe; Differential Pressure Probe; Intelligent Filter Scanning Probe; Air Velocity Probe; Spare Removable
Rechargeable Li-Ion Battery; External Battery Charger; LMS Express, RT Software; LMS Pharma Edition Software; LMS Standard Software; Validation Documentation; Printed Operating Manual; Cleanroom Cart

How - To - Videos

Solair 1100/3100/5100 Tutorial Series | Configure Sample Settings

Solair 1100/3100/5100 Tutorial Series | Configure Alarm Settings

Solair 1100/3100/5100 Tutorial Series | Configure Security Settings

Reports Function

Locations Function

Get a Quote

scanair being used by cleanroom technician
Solair - 3100
Solair - 3100
Solair - 3350
Solair - 3350
Solair-1100LD
Solair 1100LD
Solair-Boulder-Counter
Solair - BC
Solair-5100
Solair-5100

Accessories

HPD ported
HPD Ported Version
High Pressure Diffuser
HPD Vented Version
ScanAir

Unmatched Versatility

Your Solair Particle Counter Can Do It All

Besides being the best portable in the market, the Solair can be connected to a manifold and used in a real time monitoring system. Add a Scan Air Pro filter scanning probe and do filter integrity testing. High pressure gas adapters allow you to do particle testing in ultra-pure gasses. Integrate environmental sensors and you have the world’s most versatile portable particle counter.

The SOLAIR incorporates our extreme life laser diode technology sensor to produce the industry’s longest laser diode life of 20+ year MTTF (based on continuous 24/7 operation).  Using a new, lighter extended-life battery and external AC adapter, the SOLAIR 3350 is even more portable. The SOLAIR features one of the industry’s best user interfaces (UI) to configure and operate the instrument with intuitive ICON-driven screens and menus that are simple and easy to learn and use.

The SOLAIR can store up to 3,000 records of particle count data from up to 8 particle channels and the configurable recipe database can store up to 50 recipes for sampling and reports.  All data can be quickly and reliably transferred to a computer, a Facility Monitoring System or a Building Automation System.  Secure data transfer (per 21 CFR Part 11) to a USB flash drive can be accomplished by using the Lighthouse secure data transfer feature and LMS Express or Express RT software to read it. Flexible communications with these instruments is made possible by the integrated Ethernet, RS-485, USB and USB flash drive interfaces.  SOLAIR  is designed, manufactured and calibrated to ISO 21501-4 to ensure particle-counting accuracy.

Key Features:

  • 0.1 µm – 5.0 µm Particle Size Range
  • 1.0 CFM (28.3 LPM) flow rate
  • Up to 8 particle sizes
  • Number 1 in class for versatility
  • Industry’s Best User Interface
  • ISO 21501-4 Compliant
  • One of the lightest portables on the market
  • Stainless steel construction
  • 21 CFR Part 11 compliant secure data transfer via USB flash drive
  • Report wizard for cleanroom pass/fail reports: FS-209E, ISO 14644-1 (2015 version), and EU GMP Annex 1
  • Multiple communication interfaces: USB, RS-485, Ethernet, and USB flash drive
  • 5.7-inch color TFT touch screen display
  • 3,000 record data storage memory
  • 50-record configurable recipe database
  • Multiple language support:  English, French, German, Italian, Spanish, Chinese, Russian, Japanese, Korean

Key Features:

  • 0.3 µm – 25.0 µm Particle Size Range
  • 1.0 CFM (28.3 LPM) flow rate
  • Up to 8 particle sizes
  • Number 1 in class for versatility
  • Industry’s Best User Interface
  • ISO 21501-4 Compliant
  • One of the lightest portables on the market
  • 20-year MTTF laser lifespan
  • Stainless steel construction
  • 21 CFR Part 11 compliant secure data transfer via USB flash drive
  • Report wizard for cleanroom pass/fail reports: FS-209E, ISO 14644-1 (2015 version)  and EU GMP Annex 1
  • Multiple communication interfaces; USB, RS-485, Ethernet and USB flash driv.
  • 5.7-inch color TFT touch screen display
  • 3000 record data storage memory
  • 50-record configurable recipe database
  • Multiple-language support: English, French, German, Italian, Spanish, Chinese, Russian, Japanese, Korean

Key Features:

  • 0.3 µm – 25.0 µm Particle Size Range
  • 100 LPM flow rate
  • Up to 8 particle sizes
  • Number 1 in class for versatility
  • Industry’s Best User Interface
  • ISO 21501-4 Compliant
  • One of the lightest portables on the market
  • 20-year MTTF laser lifespan
  • Stainless steel construction
  • 21 CFR Part 11 compliant secure data transfer via USB flash drive
  • Report wizard for cleanroom pass/fail reports: FS-209E, ISO 14644-1 (2015 version)  and EU GMP Annex 1
  • Multiple communication interfaces; USB, RS-485, Ethernet and USB flash driv.
  • 5.7-inch color TFT touch screen display
  • 3000 record data storage memory
  • 50-record configurable recipe database
  • Multiple-language support: English, French, German, Italian, Spanish, Chinese, Russian, Japanese, Korean

Key Features:

  • 0.5 µm – 25.0 µm Particle Size Range
  • 1.0 CFM (28.3 LPM) flow rate
  • Up to 8 particle sizes
  • Number 1 in class for versatility
  • Industry’s Best User Interface
  • ISO 21501-4 Compliant
  • One of the lightest portables on the market
  • 20-year MTTF laser lifespan
  • Stainless steel construction
  • 21 CFR Part 11 compliant secure data transfer via USB flash drive
  • Report wizard for cleanroom pass/fail reports: FS-209E, ISO 14644-1 (2015 version)  and EU GMP Annex 1
  • Multiple communication interfaces; USB, RS-485, Ethernet and USB flash driv.
  • 5.7-inch color TFT touch screen display
  • 3000 record data storage memory
  • 50-record configurable recipe database
  • Multiple-language support: English, French, German, Italian, Spanish, Chinese, Russian, Japanese, Korean

Key Features:

  • 0.5 µm – 100.0 µm Particle Size Range
  • 1.0 CFM (28.3 LPM) Flow Rate
  • Simultaneous Display of up to 6 Particle Sizes
  • Number 1 in class for versatility
  • Industry’s Best User Interface
  • ISO 21501-4 Compliant
  • One of the lightest portables on the market
  • 20-year MTTF laser lifespan
  • Stainless steel construction
  • 21 CFR Part 11 compliant secure data transfer via USB flash drive
  • Report wizard for cleanroom pass/fail reports: FS-209E, ISO 14644-1 (2015 version)  and EU GMP Annex 1
  • Multiple communication interfaces; USB, RS-485, Ethernet and USB flash driv.
  • 5.7-inch color TFT touch screen display
  • 3000 record data storage memory
  • 50-record configurable recipe database
  • Multiple-language support: English, French, German, Italian, Spanish, Chinese, Russian, Japanese, Korean
  •  

Instrument Files

Title
Download
Brochure – Solair 1100LD, 3100, 3350, 5100, Boulder Counter
Download
Data Sheet – Solair 3100, 3350, 5100, Boulder Counter
Download
Operating Manual – Solair 3100, 5100, Boulder Counter Gen E
Download

Accessory Files

Title
Download
Data Sheet – High Pressure Diffuser Data Sheet
Download

HPD - Ported

Features: 

  • Allows sampling of Inert Pressurized Gasses
  • 0.1, 1.0, 2.0, 3.5 CFM Models available
  • Pressure Ranges from 5 – 150 PSI (depending on flow rate)
  • Ported & Exhausted versions available
  • Compatible with Handheld, Solair & ApexZ Instruments
  • 1.30″ (3.3 cm) at max diameter x 11.76″ [29.9 cm] max length
  • 1.50 lbs (0.68 kg)

HPD - Vented

Features: 

  • Allows sampling of Inert Pressurized Gasses
  • 0.1, 1.0, 2.0, 3.5 CFM Models available
  • Pressure Ranges from 5 – 150 PSI (depending on flow rate)
  • Ported & Exhausted versions available
  • Compatible with Handheld, Solair & ApexZ Instruments
  • 1.30 (3.3 cm) at max diameter x 10.65″ [27 cm] max length
  • 1.1 lbs (0.5 kg)

Subscribe to our Mailing List