1100 LD |
3100 |
5100 |
3350 |
Boulder Counter |
Specifications
|
Size Range |
0.1 - 1.0 μm
Optional ranges up to 5.0 µm |
0.3 - 25.0μm |
0.5 - 25.0μm |
0.3 - 25.0 µm |
5.0 - 100.0 um |
Channel Sizes |
Standard 8 channels: 0.10, 0.15, 0.2, 0.25, 0.3, 0.5, 0.7, 1.0μm
Optional 8 channels: 0.10, 0.15, 0.2, 0.25, 0.3, 0.5, 1.0, 5.0 µm |
Standard: 0.3, 0.5, 1.0, 3.0, 5.0, 10.0 µm
Standard: 0.3, 0.5, 1.0, 5.0, 10.0, 25.0 µm
8-Channel: 0.3, 0.5, 0.7, 1.0, 3.0, 5.0,7.0,10.0 µm |
Standard: 0.5, 0.7, 1.0, 3.0, 5.0, 10.0 µm
Standard: 0.5, 1.0, 3.0, 5.0, 10.0, 25.0 µm
8-Channel: 0.5, 0.7, 1.0, 2.0, 3.0, 5.0, 7.0,10.0 µm |
Standard: 0.3, 0.5, 1.0, 3.0, 5.0, 10.0 µm
Optional: 0.3, 0.5, 1.0, 5.0, 10.0, 25.0 µm |
Standard: 5.0, 10.0, 25.0, 40.0, 50.0,100.0µm |
Flow Rate |
1.0 CFM (28.3 LPM) |
100 LPM |
28.3 LPM |
Counting Efficiency |
50% @ 0.1 µm; 100% for particles > 0.15 µm per ISO 21501-4 |
50% @ 0.3 µm; 100% @ sizes
> 0.45 µm |
50% @ 0.5 um; 100% @ sizes
> 0.75 µm |
50% @ 0.3 µm; 100% @ sizes > 0.45 µm |
50% @ 5.0 um |
Concentration Limits |
500,000 Particles / ft3 (17,150,000 / m3) @ 5% Coincidence Loss |
1,000,000 particles/ft3
(35,300,000/m3) @ 10% coincidence loss
Suitable for ISO classes 1-8 cleanrooms |
400,000 particles/ft3 (14,120,000/m3)
@ 10% coincidence loss
Suitable for ISO classes 1-8 cleanrooms |
500,000 particles/ft3 @ 5% coincidence loss |
Light Source |
Long Life Laser Diode |
Extreme Life Laser Diode Technology with >20-year Mean Time To Failure (MTTF) |
Dimensions |
14.65"Lx8.0"Wx10.4"H (37x20x26cm) |
9.5"Lx8.0"Wx10.2"H (24.13x20.32x25.9cm) |
14.65"Lx8.0"Wx10.4"H (37x20x26cm) |
9.5"Lx8.0"Wx10.2"H (24.13x20.32x25.9cm) |
Weight |
With battery: 23.55 lbs (10.68 kg)
Without battery: 20.45 lbs (9.28 kg) |
With battery: 13.40 lbs (6.08 kg)
Without battery: 12.00 lbs (5.44 kg) |
With battery: 26.15 lbs (11.86 kg)
Without battery: 23.05 lbs (10.46 kg) |
With battery: 13.40 lbs (6.08 kg)
Without battery: 12.00 lbs (5.44 kg) |
Zero Count Level |
< 1 count / 5 minutes (per JIS B 9921) |
Calibration |
Meets ISO 21501-4 Calibration using NIST Traceable PSL Spheres, DMA and Condensation Particle Counter |
Count Modes |
Automatic, Manual, Beep, Concentration, Cumulative/Differential, Real Time (with LMSNet or LMS Express RT software) |
Data Storage |
3,000 Sample Records, Rotating Buffer (particle and environmental data with location and time) |
Communication Modes |
USB, Ethernet, RS-485 and Secure Data Transfer (LSDT) to and Storage on USB Flash Drive |
Supporting Software |
LMS Express and Express RT; LMS Pharma |
Environmental Sensors |
Optional (up to 4) temperature/relative humidity, air velocity, differential pressure or others |
Touch Screen Display |
5.7-inch (14.47 cm) Color TFT Touch Screen Display |
Reports |
Pass/Fail Report Wizard for ISO-14644-1, EU GMP Annex 1 and FS-209 Report Formats |
Alarms |
Integrated alarm buzzer and external alarm connector. Alarms on counts, low battery, sensor failure |
Sample Output |
Internally Filtered to 100% @ 0.1μm, Optional HEPA Filtration for Cooling System Exhaust |
Internally filtered to HEPA standards (>99.97% @ 0.3μm) |
Vacuum Source |
Internal closed-loop controlled pump. Allows the use of 3/8" ID sample tubing of up to 12 meters (40 feet) @ 50 & 100LPM |
Enclosure |
Stainless Steel |
Power |
100-240 VAC, 50-60 Hz, External AC/DC Adapter |
Battery |
Optional: Li-Ion, Removable & Rechargeable (typical recharge time 7.5 hours) |
Languages |
English (U.S.), French, German, Italian, Spanish, Chinese, Russian, Japanese, Korean |
Environmental Conditions
|
Operating |
50°F to 104°F (10°C to 40°C) / 20% to 95% non-condensing |
Storage |
14°F to 122°F (-10°C to 50°C) / Up to 98% non-condensing |
Accessories
|
Included |
Operating Manual on CD; Isokinetic Sample Probe with Tripod and Tubing; Purge Filter; Power Cable and Spare Fuse; LMS XChange and Express Software; Printer; Printer Paper |
Optional |
Printed Operating Manual; Network Adapter; Wireless Network Adapter; Internal Removeable, Rechargeable Battery; Carrying Case with Wheels; Cleanroom Cart; Temperature/Relative Humidity Probe; Air Velocity Probe; Differential Pressure Probe; Filter Scanning Probe; High Pressure Difuser; 32-Port Universal Manifold; LMS Express RT Software; LMSNet Software; Validation Documentation |
Wireless Network Adapter; Removable Rechargeable Li-Ion Battery; Carrying Case with Wheels; Temperature/Relative Humidity Probe; Differential Pressure Probe; Intelligent Filter Scanning Probe; Air Velocity Probe; Spare Removable
Rechargeable Li-Ion Battery; External Battery Charger; LMS Express, RT Software; LMS Pharma Edition Software; LMS Standard Software; Validation Documentation; Printed Operating Manual; Cleanroom Cart |